Digox 602 sodium

Measuring of the sodium value at low concentrations in steam is of great importance in power plants. This is due to the fact that sodium is an important element in the corrosive process. This measurement allows, e.g., fast monitoring of leakage in the condenser or hole in the cation exchange resin. Sodium measurement is becoming increasingly important in the revision of the VGB guidelines (S-006 / S-010).

Digox 602 sodium is an analyzer for continuous measuring of dissolved sodium, even at the level of the element trace and it is ideal, e.g., for operation in steam of power plants, for controlling full demineralization of the unit, seawater desalination and in the semiconductor and electronics industry.

Technical Specification

Device

Digox 602 sodium

Measuring range

0,1 - 9999 ppb Natrium

Measuring principle

measuring of the potential of the ion-sensitive Na-electrode against the reference electrode

Display

graphical display, measured values for each channel with time and operating conditions

Accuracy

maximum {± 2% of the reading; ± 2 ppb} within the measuring range 0.1 - 2000 ppb

maximum {± 5% of the reading; ± 5 ppb} within the measuring range of 0.1 - 9999 ppb

Repeatability

maximum {± 2% of the reading; ± 2 ppb} within the measuring range of 0.1 - 2000 ppb

maximum {± 5% of the reading; ± 5 ppb} within the measuring range of 0.1 - 9999 ppb

Calibration

Three-point calibration by a triple standard allowance

Agents

0.5 l of standard solution; 1.0 l Diisopropylamine (alkalization),  NH3 solution (option)

Data interface

Profibus DP (optional)

Binary inputs

6 control inputs for external contacts for channel suppression and remote control

Alarm outputs

two relays (1x for warnings and 1x for alarms), load at  250 VAC / 3 A, max. 24 VDC / 3 A

Operation

Password protection for two levels, entering trash values, communication parameters and measuring cycles

Analogue outputs

Up to 6 analogue outputs 4… 20 mA, max. load resistance of 500 Ω

Response time

180 seconds (95%) within the measuring range of 0.1 - 2000 ppb; 600 seconds (95%) within the measuring range of 0.1 - 9999 ppb

Ambient conditions

 +5 - +45 ° C, storage and transport 0 - 50 ° C 1), relative humidity 30 - 95%

Sample quality

10 - 40 ° C, 10 - 15 l / h, 1.0 – 5.0.bar, pH 4 - 11 (without  buffer)

Connection

1-6 input channels with an application for maintaining constant pressure, another laboratory sample possible

Power source

100 - 240 VAC 50/60 Hz, 150 VA, battery-free saving of parameters, no loss of data after power interruption

Standard compliance

Compatible with CE and CB according to EN 61010-1 / EN 61326-1 / EN 61326-2-3 / EN 50581

Protection system

IP 65 / electronic unit

Weight 30.0 kg

Dimensions

850 x 450 x 250 mm (height x width x depth)

Room for assembly

1050 x 550 x 500 mm (height x width x depth)

Main Function

  • Very low diisopropylamine consumption (conditioning): usually 0,5 l / month at pH 7 and temperature of  25 °C
  • Automatic three-point calibration,  adjustable time interval
  • Individually adjustable sequence and measurement time in a multichannel device
  • Control and monitoring of adjustable pH value
  • Usable in the range (without buffer) pH 4-11
  • Automatic temperature compensation
  • Galvanically isolated signal outputs
  • Pressure regulator and easy to clean channel pre-filter
  • Built-in sequencer for up to 6 sampling channels
  • Additional connection for measuring of the laboratory sample
  • Profibus DP interface