Digox 6.1 HY-S

Digox 6.1 is a device for measuring the concentration of hydrazine dissolved in high - purity water using amperometry. Hydrazine oxidizes on the gold electrode that is in direct contact with water. Therefore, short response times and the highest accuracy are ensured. One of the features of the Digox 6.1 model is a robust, low-maintenance sensor. Thanks to the modular design of the device, the sensor is easily accessible in the case it needs to be cleaned or repaired.

Technical Specification

Device Digox 6.1 HY-S stationary (DG 77.50.00)
Measuring range 0… 1 000 μg / l N2H4
Measuring error <1% in the measuring range of 0 ... 1 000 μg / l or ± 3 μg / l in the measuring range of 0 ... 100 μg / l
Resolution   0,1 μg / l
Response time t90 <8 s
Calbration reference calibration
Data recording 4 000 data sets in continuous saving of the data recorder with several measuring cycles possible
Flow 3… 20 l / h, compensated automatically, optimally 10 l / h
pH value stable 9 - 10 ± 0.2
Sample pressure 0… 8 bar, pressure peaks up to 16 bar
Sample temperature 0… 60 ° C, compensated automatically
Ambient temperature 0 ... 40 ° C
Signal outputs 3 x active 0 (4)… 20 mA freely adjustable; 5 x switching contact (converter), optionally 60V / 0,5A
Interface optional USB 2.0, Ethernet
Power source 100… 240 VAC (50/60 Hz), 20 VA, optionally 24 VDC (DG 77.60.00)
Protection system IP 65
Weight 8.0 kg, mounted on a stainless steel plate (optionally on a stainless steel cabinet)
Dimensions 580 x 335 x 140 mm (height x width x depth)

 

Hlavní funkce

  • Vysoká přesnost, krátká doba odezvy, žádný posun signálu
  • Robustní senzor, nízká údržba